Agilent Technologies 85050B User's And Service Manual page 32

7 mm calibration kit
Table of Contents

Advertisement

Use, Maintenance, and Care of the Devices
Gaging Connectors
desired value. The pin depth of the sliding load is preset at the factory. The pin depth should not
have to be reset each time the sliding load is used, but it should be checked before each use.
If the pin depth is outside the observed limits listed in
procedure to reset it. Always measure the sliding load pin depth before attaching it to any
connector.
This procedure assumes that you were directed here from
If not, perform the steps in that procedure before performing this procedure.
1. The gage should be attached to the sliding load. Refer to
if necessary.
With a 0.050 inch hex key, loosen the two largest hex screws by turning them 1/4 turn clockwise.
2.
Refer to
Figure
3-5.
Gently turn the knurled center conductor cap on the sliding load until the gage pointer reads
3.
0.0001 inches. Refer to
Tighten the two hex screws just until they are finger tight (do not overtighten).
4.
Wait approximately five minutes to allow the temperature to stabilize. Do not touch either the
5.
gage or the sliding load during this time.
Note the gage reading. If it is no longer within the allowable range, perform steps 2
6.
Remove the gage from the sliding load. Replace the center conductor tip in the sliding load and
7.
remove the aligning pin from the gage.
The sliding load pin depth is now in specification and the load is ready to use. Once the sliding load
pin depth is set it rarely needs to be adjusted. However, the pin depth should be rechecked before
each use. Replace the protective plastic caps on the sliding load and gage connectors when these
devices are not in use.
Figure 3-5 Adjusting the Sliding Load Pin Depth
3- 12
Figure
3-5.
Table 2-2 on page
2-5, use the following
"Gaging the Sliding Loads" on page
"Gaging the Sliding Loads" on page 3-11
3-11.
5 again.
85050B