GE EPS 3501 XL User Manual page 13

Electrophoresis power supply
Hide thumbs Also See for EPS 3501 XL:
Table of Contents

Advertisement

1. Choose Program Number
(1
2)
2. Choose Step or Gradient
mode. (Step)
3. Confirm Phase number (1)
4. Set Voltage limit (step) or
Voltage endpoint
(gradient).
(0
1000 V)
5. Set Current limit.
(0
10 mA)
6. Set Power limit.
(0
10 W)
7. Choose Breakpoint unit
(h, Vh or mAh) or no
breakpoint (OFF).
(h
Vh)
(h
OFF)
8. Set Breakpoint
(0
1000 Vh)
9. Last Phase,
YES or NO?
10. Choose Alarm OFF or ON
(ON).
11. The program is ready.
Fig 3. Step-by-step summary of programming.
Electrophoresis Power Supply EPS 3501 XL User Manual 18-1130-22 Edition AD
1 ⁄ 1
SET:
1 Vh
1 mA
5 W
500 V
Select by
Confirm by
SET
2 1
STEP/gradient
Confirm by
Select by
2 1
SET:
0:00 h
0 V
0 mA
0 W
set
Confirm by
enter
0:00 h
SET:
2 1
0 mA
0 W
0 V
Select by
Confirm by
0:00 h
2 1
SET:
1000 V
10 mA
0 W
Select by
Confirm by
0:00 h
SET:
2 1
1000 V
10 mA
0 W
Confirm by
Select by
1st Phase
Other phases
2 1
0:00 h
SET:
1000 V
10 mA
10 W
Select by
Confirm by
Vh
0 Vh
SET:
2 1
1000 V
10 mA
10 W
Select by
Confirm by
1000 Vh
SET:
2 1
Last phase
?
Y
Select by
Confirm by
YES
Alarm Phase 1?
N
Phase: - - - -
Select by
Confirm by
2 1
A
1000 Vh
SET:
1000 V
10 mA
10 W
set
enter
set
enter
set
enter
set
enter
set
enter
set
Breakpoint OFF
enter
set
enter
set
enter
NO
set
enter
2 1
SET:
1000 V
10 mA
Operation 5
A
1000 Vh
10 W
13

Advertisement

Table of Contents
loading

Table of Contents