SERVICE GUIDE

Find PM Gain (Gen Rad)

Find PM Gain (Portals—RT)
[8] Click [OK] and insert the Cassette which has been exposed with the required dose into the Scanner.
Note
When the calibration process has been completed successfully, a green check mark appears next to the PM nominal
calibration field along with the date of the calibration procedure.

Calibration Successful Screen

Note
If the calibration was not successful, a red X appears next to the calibration field.

Calibration Failed

[9] Make sure the Scanner is not located near a window or other strong light source and repeat the calibration
process.
[10] If the calibration still fails, contact Kodak support.
74
10APR2006 – Catalog No. 6H1344