Test; Calibrations; Checks - Siemens MAMMOMAT 1000 Maintenance Instruction

Table of Contents

Advertisement

Testing
4
Testing
4.1

Test

PMF
Test of units
For instructions how to test the units see Service Instructions for Opdima
units .
4.2

Calibrations

PMF
CCD-Camera calibration
For instructions how to calibrate the CCD-Camera see Service Instructions for Opdima
chapter CCD camera calibration and maintenance .
Q
Calibration of the biopsy unit
Check biopsy accuracy according to Service Instructions for Opdima
the calibration of the biopsy unit . Verify the calibration for all four target points of the
stereo calibration phantom.
If biopsy accuracy is not satisfactory, perform calibration according to Service
Instructions for Opdima r , chapter Calibration of the biopsy unit .
4.3

Checks

QIQ
Opdima
r
For instructions how to check Opdimar AEC see Service Instructions for Opdima
chapter Check of Opdima AEC .
QIQ
Resolution
For instructions how to check resolution see Service Instructions for Opdima
Check of resolution .
Siemens AG
Medical Solutions
AEC
Register 8
SPB7-230.101.03
Rev. 04
Page 1 of 2
07.03
CS PS 24
4 - 1
r
, chapter Test of
r
,
r
, chapter Verifying
r
,
r
, chapter
Opdimar
Maintenance Instructions

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mammomat 3000Mammomat 3000 nova - opdima

Table of Contents