Testing
4
Testing
4.1
Test
PMF
Test of units
For instructions how to test the units see Service Instructions for Opdima
units .
4.2
Calibrations
PMF
CCD-Camera calibration
For instructions how to calibrate the CCD-Camera see Service Instructions for Opdima
chapter CCD camera calibration and maintenance .
Q
Calibration of the biopsy unit
•
Check biopsy accuracy according to Service Instructions for Opdima
the calibration of the biopsy unit . Verify the calibration for all four target points of the
stereo calibration phantom.
•
If biopsy accuracy is not satisfactory, perform calibration according to Service
Instructions for Opdima r , chapter Calibration of the biopsy unit .
4.3
Checks
QIQ
Opdima
r
For instructions how to check Opdimar AEC see Service Instructions for Opdima
chapter Check of Opdima AEC .
QIQ
Resolution
For instructions how to check resolution see Service Instructions for Opdima
Check of resolution .