EMI Detector Test

Test Limits

Test Description

EMI Detector
Procedures
200 Hz Procedure
9 kHz Procedure
120 kHz Procedure
Test Description
The internal 50 MHz signal is applied to the input of the analyzer. For each of the CISPR
filters (200 Hz, 9 kHz, and 120 kHz) a peak, quasi-peak, and average measurement is
made on the 50 MHz CW signal.
Item
Adapters
Type-N (m) to BNC (f)
Cables
BNC (m) to BNC (m),
9 inches
Figure 4-3. Equipment Setup
Chapter 4
Test Limits)
Peak, Quasi-Peak, or Average
Values
+/– 2 dB
Critical Specifications
(for this test)
ESA Functional Tests

EMI Detector Test

Recommended
Agilent Model
1250-0780
HP 10502A
159