P5514, P5514-V & P5514-EP
Users Manual
1
Connect the Gauge
Refer to Figure 1 for steps 1 to 4.
1. Screw the appropriate gauge adapter fully on to the instrument to be tested.
2. Screw assembly down COUNTER-CLOCKWISE on to test port.
Hand-tight is sufficient; ensure that the bottom face contacts the O-ring on
the test port.
3. To adjust the position to face forward, hold the gauge adapter and turn the instrument
4. Hold the instrument steady, while turning the gauge adapter
Connect the Test Port Insert
For devices with 1/8 BSP or NPT mounting threads, the diameter of the thread is very
close to the effective sealing diameter of the O-ring fitted to the test port.
This can make it difficult to achieve a good seal. When mounting these devices, use the
test port insert (stored in the spare seals container) as shown Figure 2.
4
2
Figure 1. Connecting the Gauge
COUNTER-CLOCKWISE, so that it faces forward.
COUNTER-CLOCKWISE until it pulls down onto the O-ring.
3
Note
4