Chapter 3 Features and Functions
Calibration Overview
Calibration Message
The function generator allows you to store one message in calibration
memory in the mainframe. For example, you can store such information
as the date when the last calibration was performed, the date when the
next calibration is due, the function generator's serial number, or the
name and phone number of the person to contact for a new calibration.
• You can record a calibration message only from the remote interface
and only when the function generator is unsecured. You can read the
message from either the front-panel or over the remote interface.
You can read the calibration message whether the function generator
is secured or unsecured.
• The calibration message may contain up to 40 characters (additional
characters are truncated).
• Storing a calibration message will overwrite any message previously
stored in memory.
• The calibration message is stored in non-volatile memory and does not
change when power has been off or after a remote interface reset.
• Front-Panel Operation: Press
softkey from the "Test / Cal" menu. The calibration message is listed
as one of the messages on the display.
• Remote Interface Operation: To store the calibration message, send the
following command, including any desired string within the single
quotes (' ').
CAL:STR 'Cal Due: 01 August 2008'