3 Sample Sessions: Measuring a Trace
3 Cursor to End Thres. and press S
to change a numerical setting" on page 132 to select a
new threshold value.
If you select an end threshold of, for example, 3.0 dB,
an End will be set at the first event with an insertion
loss of 3 dB or more. If you select a threshold of 0 dB,
no End will be set.
4 Select [A
]<S
NALYSIS
CAN
The first Event which exceeds the specified End
Threshold is now set to type End, and subsequent
Events are ignored.
Or
1 Use the Cursors to move the current marker to an
Event.
Select [E
]<D
VENTS
ECLARE
An End is set at the event near the current marker.
If the current marker is not at an event, no End is set.
The end Event is listed as type End in the Event table, and
marked on the Event Bar with a special symbol (see
Figure 32). All events to the right of the End Event are
removed
. Follow "How
ELECT
T
> to run a scan trace.
RACE
E
> from the popup panel.
ND
99