Output; Ttltrg [:State] - HP E1326B User Manual

5 1/2 digit multimeter
Table of Contents

Advertisement

Subsystem Syntax
n
:TTLTrg
[:STATe]
Parameters
Comments
153
Command Reference
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The

OUTPut

command subsystem enables you to route the multimeter's voltmeter
complete signal to the VXIbus TTL trigger lines.
OUTPut
:TTLTrgn[:STATe] <mode>
:TTLTrgn[:STATe]?
n
OUTPut:TTLTrg
[:STATe] <
complete signal to the specified VXIbus trigger line (TTLTrg0 through TTLTrg7)
on the backplane P2 connector.
Parameter Name
n
<
mode
>
The
OUTPut
subsystem applies to the HP E1411B multimeter only.
You can substitute decimal values for the
The voltmeter complete signal is always routed to the E1411B multimeter's front
panel "VM Complete" BNC connector. When enabled (
command also routes voltmeter complete to the specified trigger line on connector
P2. When disabled (
OFF
front panel connector.
The multimeter generates the voltmeter complete signal after it has sampled the input
for each reading. The length of time this low-going TTL signal is true (low) depends
on the aperture time and on the autozero mode as shown below.
Aperture Time
320 ms (50 Hz)
267 ms (60 Hz)
20 ms (50 Hz)
16.7 ms (60 Hz)
2.5 ms (400 Hz)
The VXIbus trigger lines are open-collector TTL lines that remain in a
non-asserted (high) state until the voltmeter complete signal is sent.
*RST Condition:
OUTP:TTLT
mode
enables or disables routing of the voltmeter
>
Parameter Type
Range of Values
discrete
0 | 1 | 2 | 3 | 4 | 5 | 6 | 7
boolean
OFF | 0 | ON | 1
OFF
("
0
") and
), voltmeter complete is routed only to the multimeter's
Autozero ON
350 ms
370 µs
20.5 ms
17.2 ms
3.1 ms
100 µs
520 µs
10 µs
n
OFF
OUTPut
Default Units
none
none
ON
("1") parameters.
), the
ON
OUTPut
Voltmeter Complete low
Autozero OFF
350 µs
370 µs
370 µs
390 µs
430 µs
250 µs
NA
70 µs
Chapter 5

Advertisement

Table of Contents
loading

This manual is also suitable for:

E1411b