Example

:ZERO:AUTO

Parameters
Comments
Example
:ZERO:AUTO?
Chapter 5
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Querying the Line Reference Frequency
CAL:LFR 50
CAL:LFR?
enter statement
CALibration:ZERO:AUTO <
voltage and resistance measurements.
Parameter Name
mode
<
>
You can substitute decimal values for the
When autozero is
, the multimeter makes a zero measurement (measurement
ON
with input disabled) following every measured reading and subtracts the zero
measurement from the reading. This doubles the time required per reading.
When autozero is
, the multimeter makes one zero measurement and subtracts
OFF
this from all subsequent measurements. A new zero measurement is made
whenever the function is changed. Notice that the mode parameters
ONCE
have the same effect.
An autozero measurement is not made following a range change whether autozero
is
or
.
ON
OFF
Autozero cannot be used when a 10 µs aperture time is specified.
The
and
CONFigure
MEASure
*RST Condition:
CAL:ZERO:AUTO ON
Turning Autozero OFF
CAL:ZERO:AUTO OFF
CALibration:ZERO:AUTO?
is enabled or disabled: "
1
" =
The number is sent to the output buffer.
!Reference frequency is 50 Hz.
!Query for reference frequency.
!Enter value into computer.
mode
enables or disables the autozero mode for DC
>
Parameter Type
Range of Values
boolean
OFF | 0 | ON | 1 | ONCE
OFF
("0") and
commands turn autozero
!Disable autozero.
returns a number to show whether the autozero mode
ON
, "
0
" =
OFF
or
ONCE
Command Reference
Default Units
none
ON
("1") parameters.
OFF
and
.
ON
.
124