Setting Emissivity
This manual will outline three ways to set the emissivity of the SP-001 / SP-002:
•
Known Emissivity Value
•
Setting Emissivity Value via Emissivity Table
•
Setting Emissivity via Secondary Temperature Device
Note:
The default factory setting for emissivity is set to 1.
Known Emissivity Value
If the emissivity of the target material is already known, you may enter it in the Emissivity Dropdown and click Apply
Settings to finalize changes. See Figure 11.
Setting Emissivity via Emissivity Table
For a list of reference materials and their corresponding emissivity values, refer to the Emissivity Table in Appendix A. Enter
the corresponding value in the Emissivity Dropdown and click Apply Settings to finalize. See Figure 11.
Setting Emissivity via Secondary Temperature Device
Note:
Ensure the SP-001 / SP-002 is connected to Omega Sync and pointed at the material being measured at a
distance that covers the entire field of view.
Step 1:
Click Calibration. See Figure 12.
Step 2:
Use a secondary device to measure the temperature of the target material being measured by the SP-001 / SP-
002.
Tip:
Recommended devices to measure the temperature of the target material include a thermocouple, RTD, etc.
Figure 11
Figure 12