Removal Of Rear Castor Arm - Sunrise Medical Quickie Q700M Service Instructions Manual

Table of Contents

Advertisement

Removal of rear castor arm

Preparation
Note! Switch off the wheelchair via remote control
Instructions
Note! Suitable work area required!
• Follow the illustrated instruction for removing the rear
castor arm from the wheelchair base.
• Rebuild the wheelchair in reverse order once completed.
• In order to complete this, you must first tilt the chair
back, electircally disconnect the batteries and remove the
batteries.
Fig. 6.2.25
1. Unscrew and remove bolt (A) from the suspension spring
Tightening torque for (B) bolts is 12Nm.
Fig. 6.2.26
2. Unscrew and remove bolt (B) bolt from the rear castor arm
Tightening torque for (B) bolts is 12Nm.
Fig. 6.2.27
3. Remove the connecting cable (C) from the underside of the rear
castor arm, which connects the wheelchair base
26
Tools used
6
5
1
A
2
B
19
Fig. 6.2.25
Fig. 6.2.26
Fig. 6.2.27
3
C
Q700M Rev.A

Advertisement

Table of Contents
loading

This manual is also suitable for:

Quickie q700-up mQuickie q700-up f

Table of Contents