Reagents And Calibration Standards; Reagents - Hach BioTector B3500dw User Manual

Online toc
Table of Contents

Advertisement

Section 6
6.1

Reagents

BioTector TOC analyzer uses the following reagents:
I. Acid: 6.0 N Sulfuric Acid (H
II. Base: 4.0 N Sodium Hydroxide (NaOH) Reagent
Reagents should not contain high levels of organics, nitrates and phosphates. Ideally, the level of organics,
nitrates and phosphates should be less than 100 g/l (ppb) in the deionized water used to prepare TOC
analyzer reagents.
Acid and Base reagents are stable up to 1 year. The reagents must be stored in a safe and secure location,
where the temperature does not drop below 2°C, in compliance with the local regulations. Table 6 below
summarizes the total days each BioTector TOC Analyzer reagent lasts at various system configurations:
Table 6 BioTector TOC Analyzer Reagent Consumption
Container Size
REAGENTS
Acid
Base
Above table is derived from several online operation parameters such as 100% online time.
Recommended bunds (reagent spill trays) to contain above quantity reagents are 1x 50 Liters.

Reagents and Calibration Standards

SO
) Reagent containing 350 mg/l Manganese Sulfate Monohydrate
2
4
(Liters)
19
19
TOTAL DAYS REAGENT LASTS
239
239
Page 64