Hach BioTector B3500ul User Manual page 95

Online toc analyzer
Table of Contents

Advertisement

Engineering mode (VOC Analysis):
TIME
The time the reaction started.
DATE
The date the reaction started.
S1:2
Reaction type and range.
CO2z
CO
analyzer zero adjust for the current reaction.
2
TCµgu
The un-calibrated TC value in µgC/l (measured value from the TC analysis).
TCµgc
The calibrated TC value in µgC/l (measured value from the TC analysis).
CO2p
The height of the TC CO
TICµgu
The un-calibrated TIC value in µgC/l (measured value from the TIC&TOC analysis).
TICµgc
The calibrated TIC value in µgC/l (measured value from the TIC&TOC analysis).
CO2p
The height of the TIC CO
NPOCµgu
The un-calibrated NPOC value in µgC/l (measured value from the TIC&TOC analysis).
NPOCµgc
The calibrated NPOC value in µgC/l (measured value from the TIC&TOC analysis).
CO2p
The height of the NPOC CO
The calculated TOC value in µgC/l (TOCµgc is calculated as TCµgc – TICµgc).
TOCµgc
The calculated VOC value in µgC/l (VOCµgc is calculated as TCµgc – TICµgc – NPOCµgc).
VOCµgc
COD/BODµgc
The calculated COD or BOD value in µgO/l (if activated in COD/BOD Program menu).
BT_DegC
BioTector temperature in Degrees Celsius (°C).
MB_DegC
Motherboard temperature in Degrees Celsius (°C).
Atm
Atmospheric pressure in kPa.
SAMPLE
Sample quality (%) from Sample Sensor signal, which is used to activate SAMPLE STATUS output.
SMPL PUMP
The five items, which are number coded or a number data, gives information on the Sample Pump:
1) operation mode (0=time mode or 1= pulse mode),
2) number of pulses during operation such as injection,
3) total time (milliseconds) taken for total number of pulse operation (see point 2 above),
4) the time (milliseconds) taken for the last pulse operation (see point 2 above),
5) error counter (ranges from 0 to 6). When a pulse is missed or not detected, the pump switches into
time mode for that specific operation (for instance, injection, synchronization, etc). System only
generates a pump warning and logs into fault archive if there are 6 consecutive failures.
ACID PUMP
Similar information on Acid Pump operation (see items 2, 3 and 4 only, listed for SMPL PUMP above).
BASE PUMP
Similar information on Base Pump operation (see items 2, 3 and 4 only, listed for SMPL PUMP above).
The status of the Cooler. When the Cooler is off, "OFF" is printed in the reaction archive data.
COOLER
The status of the Ozone Destructor Heater. When the heater is off, "OFF" is printed in the data.
O3 HEATER
Note that the COD or BOD result is added into the reaction screens and reaction archives if the COD or BOD
is activated in COD/BOD Program menu.
peak.
2
peak.
2
peak.
2
Page 95