Supplements to Service Instructions
Recalibration of the DAP measuring system
This instruction shows how to recalibrate the DAP measuring system by using an external
ionization chamber (dose meter).
Required equipment
•
Dose meter
•
Calibration tool for DAP measuring system, Part No. 65 84 978
Recalibration procedure
1.
System OFF.
2.
Disconnect the DAP ionization chamber connector which is connected to board
D70 (pos. 1 / Fig. 1).
3.
Connect the DAP calibration tool between the connector and board D70.
NOTE
4.
System ON.
Siemens AG
Medical Solutions
Fig. 1 Connector
It is irrelevant which connector is the input or the output of the
DAP calibration tool.
SPR8-220.031.04
Rev. 04
1
Page 1 of 4
09.03
CS PS 24
3 - 1
3
3
3
3
MOBILETT Plus/ -E/ -HP