Recalibration Of The Dap Measuring System; Required Equipment; Recalibration Procedure - Siemens MOBILETT Plus Installation Instructions Manual

Hide thumbs Also See for MOBILETT Plus:
Table of Contents

Advertisement

Supplements to Service Instructions

Recalibration of the DAP measuring system

This instruction shows how to recalibrate the DAP measuring system by using an external
ionization chamber (dose meter).

Required equipment

Dose meter
Calibration tool for DAP measuring system, Part No. 65 84 978

Recalibration procedure

1.
System OFF.
2.
Disconnect the DAP ionization chamber connector which is connected to board
D70 (pos. 1 / Fig. 1).
3.
Connect the DAP calibration tool between the connector and board D70.
NOTE
4.
System ON.
Siemens AG
Medical Solutions
Fig. 1 Connector
It is irrelevant which connector is the input or the output of the
DAP calibration tool.
SPR8-220.031.04
Rev. 04
1
Page 1 of 4
09.03
CS PS 24
3 - 1
3
3
3
3
MOBILETT Plus/ -E/ -HP