Siemens MOBILETT Plus Installation Instructions Manual page 15

Hide thumbs Also See for MOBILETT Plus:
Table of Contents

Advertisement

Installation
8.
NOTICE
Pull out the cables (pos. 1/ Fig. 8) and make a loop. Wrap the cables with pin plugs
K21 and K22 into the spiral cable tie (included in the delivery). Attach the spiral
cable tie from approximately 2 cm above the bar (Fig. 8) and place it between the
two thick cables.
2
9.
Place a cable tie (4.7x195mm) around the spiral cable tie (pos. 2 / Fig. 8).
Siemens AG
Medical Solutions
For MOBILETTs with remote exposure switch system
installed:
Mount the sensor on the new lower arm cover and
thread the cable as previously described. Let the cover
hang on the side of the arm (see pos. 6 / Fig. 8). If there
are stickers on the old cover, transfer them to the new
cover.
Wrap the remote control cable into the spiral cable tie
together with the other cables according to step 8. by
alternatingly turning it and pulling it downwards.
1
Fig. 8 Insertion of cables into the MOBILETT arm.
SPR8-220.031.04
Rev. 04
5
4
Page 7 of 30
09.03
CS PS 24
2 - 7
6
3
MOBILETT Plus/ -E/ -HP

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mobilett plus eMobilett plus hp

Table of Contents