8.
Disconnect pin plug K52 from board D8 (pos. 1 / Fig. 24).
9.
Fasten the cable shielding connector with the screw holding the protective cover
over board D8 (pos. 4 / Fig. 24). Replace the previously mounted washer for a
washer (KB 5.1mm) included in the kit.
10.
Attach pin plug K52b of the cable to pin plug K52, then attach pin plug K52a to K52
on the board (pos. 1 / Fig. 24).
11.
Thread the ends with pin plugs K24 and K25 through the upper hole (pos. 2 / Fig.
24) and through the lower hole (pos. 3 / Fig. 24).
Fig. 24 Threading cable to new power supply
MOBILETT Plus/ -E/ -HP
D8
SPR8-220.031.04
Rev. 04
2
3
Page 20 of 30
09.03
CS PS 24
Installation
4
1
Siemens AG