114 | DR 14s | Remarks for HF-emission and immunity
Magnetic field
at the supply
frequency
(50/60 Hz) in
accordance
with IEC
61000-4-8
GSM modula-
tion
ENV 50204
Radiated RF
IEC 61000-4-3
Tests of Resist-
ance to Disrup-
tion
0350B EN 20190319 1349 *
3 A/m
3 V/m
900 MHz
modulated @
200 Hz (square
signal)
3 V/m
80 MHz to 2.5
GHz
80 MHz to 1
GHz
Note: Ur is the alternating voltage.
IEC 60601
Level of
Test Level
Agree-
ment
3 A/m
Magnetic field at the net-
work frequency should
correspond to the typical
values as they are in a
commercial and clinical
environment.
3 V/m
Interference may occur
in the vicinity of equip-
900 MHz
ment marked with the
modulated @
following symbol:
200 Hz (square
signal)
3 V/m
Interference may occur
in the vicinity of equip-
80 MHz to 2.5
ment marked with the
GHz
following symbol:
80 MHz to 1
GHz
Electromagnetic Environment
Use portable and mobile radio
sets at a safe distance from the
DR Detector (including the
leads) not closer than the rec-
ommended protective distance,
which is calculated according to
the equation suitable for the
transmission frequency.