Qc Test Image - AGFA Drystar AXYS XL User Manual

Hide thumbs Also See for Drystar AXYS XL:
Table of Contents

Advertisement

QC test image

The QC image has several additional fields where the QC data can be filled in.
This image should be filed as part of the QC procedure.
The following two examples illustrate a QC test image for a different film
format.
• QC test image for film format 14x17"
Quality Control Test Image
• QC test image for film format 8x10"
Drystar AXYS XL | Advanced Operation (key-operator mode) | 95
Note: The dimensions A and B are not indicated on the printed
film. The QC objects on printed films may look different than in
the illustration. The location, size and shape of the QC objects
do not jeopardize the functionality of the QC and can be altered
depending on software versions, film types and film sizes.
1
2
3
Identification
Initials:
Date:
Time:
Geometry test
Daily density tests
Max D
Hi D
Mid D
Lo D
Base
+ Fog
DD
Density Difference
(
H i D - L o D )
Weekly spatial
Resolution test
Sets of dots visible?
Yes
No
3752B EN 20210601 1655