Charts for general radiography QC
Imager Type: __________
Film Type: ____________
Step 1: Print QC Test images on five consecutive days. Record the optical densities
measurements in the tables below. After five days, average the values to determine
the reference levels for each of the parameters.
Month
Day
Initials
Low Density
Mid Density
High Density
Average of 5 Values = calculated reference Hi g h D e nsi ty l e v el
Step 2: Copy the calculated reference levels to Charts 2A/B ('Daily Density Control Chart')
Figure 11: Chart 1, Determination of Operating Levels
Determination of Operating Levels
Serial #: _________________
Emulsion #: ______________
Day 2
Average of 5 Values = calculated reference L ow D e nsi ty l e v el
Average of 5 Values = calculated reference "Mid Density" level
Drystar AXYS XL | Quality Control Charts | 139
Date ____________________
Day 4
Day 5
3752B EN 20210601 1655