Safety

Safety notes

The following applies in potentially explosive environments:
- For vibration measurements, you may only use
>
LineDrive transducers of the VIB 6.1xx DEX model series
>
VIBCODE transducers (VIB 8.660 HEX)
- Under no circumstances may you take measurements with the
following sensors:
>
TIPTECTOR hand-held probe,
>
LineDrive transducers of the VIB 6.1xx EX model series
>
LineDrive transducers of the VIB 6.1xx REX model series.
- The connection cable for extra-low signal current (VIB 5.434) and
extra-low signal voltage (VIB 5.433) may not be used with VIBX-
PERT EX.
- Under no circumstances should you recharge the battery.
- Under no circumstances should you use the VIBXPERT case (VIB
5.329 X).
- The protective film must be removed from the display.
How to deal with rechargeable batteries
- Charge the battery only outside the EX area.
- Do not send devices with defective battery by air freight.
- Defective batteries may only be replaced by authorized
PRÜFTECHNIK personnel.
Also observe the safety notes in the operating manual.

Conformity

The product complies with the relevant European directives. The
declaration of conformity is available as a PDF and may be down-
loaded from the PRÜFTECHNIK homepage at:
www.pruftechnik.com/com/Know-How/Media-Center/
Certificates-and-Material-Saftey