Agilent Technologies 8960 Series Reference Manual page 301

Hide thumbs Also See for 8960 Series:
Table of Contents

Advertisement

SETup:WBERror:CONTinuous
Function
Sets/queries the trigger arm state for the loopback bit error ratio measurement. The trigger arm
state determines whether the bit error measurement will be made once then stop (0|OFF
(single)), or automatically re-arm upon completion of one measurement and repeat the process
(1|ON (continuous)).
Setting
Range: 1|ON|0|OFF
Query
Range: 1|0
*RST Setting
0 (off)
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"SETUP:WBERROR:CONTINUOUS ON
!Sets measurement to single trigger mode.
SETup:WBERror:COUNt
Function
Sets/queries the number of bits to test in the loopback bit error ratio test.
Setting
Range: 1000 to 999000
Query
Range: 1000 to 999000
*RST Setting
10000
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"SETUP:WBERROR:COUNT 5000"
SETup:WBERror:TIMeout[:STIMe]
Function
The setting form of this command performs two functions: (1) sets the loopback bit error ratio
measurement's timeout (see
leakage measurement's timeout STATe to 1 (on).
The query form returns the loopback bit error rate measurement's timeout value.
Setting
Range: 0.1 to 999.9 seconds
Resolution: 0.1 second
Query
Range: 0.1 to 999.9
*RST Setting
20 seconds, on
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"SETUP:WBERROR:TIMEOUT:STIME 5 S"
!Sets timeout state to ON and sets the timeout time to 5 seconds.
"