Instruction Manual
WD290
Article number: 10268-0068
Version: 001-03/08

Advertisement

Table of Contents
loading

Summary of Contents for BELIMED WD290

  • Page 1 Instruction Manual WD290 Article number: 10268-0068 Version: 001-03/08...
  • Page 2: Table Of Contents

    Washing and disinfection ......22 2/39 Instruction Manual WD290 10268 © Belimed...
  • Page 3 Glossary..........36 Organisation Belimed AG ......38 20.1...
  • Page 4: Introduction

    There may be a risk to the life and health of persons. CAUTION! A situation for which there is a warning of damage to property and equip- ment. NOTE User tips and useful information on the best possible use of the equipment. 4/39 Instruction Manual WD290 10268 © Belimed...
  • Page 5: For Your Safety

    Validation of the programme parameters must be performed regularly (see Chap. 2.6 "Process validation") • Installation, deinstallation, servicing or modification must only be carried out by persons authorised by Belimed • In the case ofincorrectly installed, operated or maintained devices all warranty claims are invalidated Non-intended use All other applications are considered as non-intended use.
  • Page 6: Fields Of Application For The Device

    Routine testing / Annual requalification (refer to appendix EN ISO 15883-1 November 2001, Chap. 6, Pages 35-37) NOTE Further information on process validation may be obtained from Belimed Customer Service. CAUTION! Validation must only be carried out by authorised persons!
  • Page 7: Device Description

    Wash compartment doors made of double-layer safety glass Operating panel Main switch (with EMERGENCY OFF function) Service panel Location type plate Transfer trolley Rack Automatic wash compartment doors 10 Door safety switching strip 7/39 Instruction Manual WD290 © Belimed 10268...
  • Page 8: Device Clean Side (Cs)

    Device description Device clean side (CS) Abb 301 Door drive (behind the panel) Operating panel with door button Wash compartment doors made of double-layer safety glass Service panel 8/39 Instruction Manual WD290 10268 © Belimed...
  • Page 9: Controller Unclean Side (Us)

    Device description Controller unclean side (US) 1) Display – With screensaver "BELIMED INFEC- TION CONTROL"; this means that this display automatically appears af- ter approx. 1 h. Press any button, dis- play"Program ready" appears again Program 1 2) Cursor left –...
  • Page 10: Pre-Treating Medical Devices

    • Necessary knowledge of the medical devices to be reconditioned • Maintenance plan and regular inspections Belimed recommends producing work instructions which describe the procedure within a working process. Preparing medical devices CAUTION! Not all medical devices are suitable for mechanical reconditioning (see Chap.
  • Page 11: Pre-Cleaning Pre-Treatment

    Pouring antiseptic solutions on the items to be rinsed • Aldehyde and alcohol vapours • Heat pre-treatment Small quantities of foam are permissible with manual cleaning in the immersion bath or ultrasonic cleaning. 11/39 Instruction Manual WD290 © Belimed 10268...
  • Page 12: Preparing The Device

    Fault without process interruption Rectify fault (see Chap. 16 "Faults") Example display text: Dosing device Empty Fault with process interruption Report to the technical specialist Example display text: No pressure Fault Code 110 12/39 Instruction Manual WD290 10268 © Belimed...
  • Page 13: Self-Disinfection

    Desinf A068 89°C Legend: SD |_| = program step status 13min = remaining run time Desinf = program name A068 = current A value 89°C = temperature of rinsing agent or drying air 13/39 Instruction Manual WD290 © Belimed 10268...
  • Page 14 Program name Correctly Finished NOTE For manual loading/unloading, unload the rack back to the US again at the end of the program. The CS door remains locked. Self-disinfection is inactive for automatic loading/unloading. 14/39 Instruction Manual WD290 10268 © Belimed...
  • Page 15: User Identification

    For an input of No. 0 or 66 "No identification" is output. Example 1: Name M. Smith H. Jones No Identification No Identification Device ready Display text: Program ready Display text for a prolonged period of disuse (see Chap. 6 "Self-disinfection"): Self Disinfection Start 15/39 Instruction Manual WD290 © Belimed 10268...
  • Page 16: Identification Via Barcode Reader

    First name / last name Heike Meyer No Identification 0 / 66 Program ready Display text: Program ready --.--.---- --:-- Display text for a prolonged period of disuse (see Chap. 6 "Self-disinfection"): Self Disinfection Start 16/39 Instruction Manual WD290 10268 © Belimed...
  • Page 17: Loading And Identifying Racks

    The length of the name input is limited to 12 characters. If the no. 0 or 66 is read"No identification" is output (see Chap. 7.2 "Identification via barcode reader") Barcodes can be produced oneself from the Internet (free of charge). (http:// www.barcodemagic.com/barcodemagic.html). 17/39 Instruction Manual WD290 © Belimed 10268...
  • Page 18: Identification Of Batch Content

    The length of the name input is limited to 12 characters. If the no. 0 or 66 is read "No identification" is output (see Chap. 7.2 "Identification via barcode reader") Barcodes can be produced oneself from the Internet (free of charge). (http:// www.barcodemagic.com/barcodemagic.html). Example: Abb 227 18/39 Instruction Manual WD290 10268 © Belimed...
  • Page 19: Loading From The Unclean Side

    Sharp or pointed objects may fall down when sliding the rack. This can cause injuries. The rinsing arms must not be blocked by too high objects or objects protruding downwards. Check rotation by hand! 19/39 Instruction Manual WD290 © Belimed 10268...
  • Page 20 The device is ready to start. Program ready dd.mm.20yy hh:mm NOTE Check that the date and time are correct. This is important for the batch doc- umentation. Report discrepancies to the technical specialist! 20/39 Instruction Manual WD290 10268 © Belimed...
  • Page 21: Automatic Rack Station (Optional)

    For correct and intended operation and servicing it is absolutely necessary to fol- low the instructions supplied wit the automated rack station. Abb 206 Rack Rack station Emergency Stop Rack station operating buttons Automatically driven rollers Rack station 21/39 Instruction Manual WD290 © Belimed 10268...
  • Page 22: Washing And Disinfection

    If smoke is emitted or water escapes, immediately turn off at the main switch and isolate the device from the mains supply. Inform the technical specialist. The end of the program is indicated with an acoustic signal Display text: Program name Correctly Finished 22/39 Instruction Manual WD290 10268 © Belimed...
  • Page 23: Unloading From The Clean Side

    CAUTION! There must be no more soiling (incrustations, depositions) visible. Detach the transfer trolley with the lever (Fig. 204 Pos. 3) Washed items ready for further processing 23/39 Instruction Manual WD290 © Belimed 10268...
  • Page 24: Switching Off The Device

    If the device is switched off at the main switch, the controller is also no longer active. Consequently, the water level is not checked and, in the case of a bro- ken valve, water can overflow. 24/39 Instruction Manual WD290 10268 © Belimed...
  • Page 25: Daily Servicing And Cleaning Work

    Twist the clamping fixture (Fig. 177 Pos. 1) counter-clockwise and pull it up- wards Disassemble the remaining components acc. to Fig. 177 Check the seal on the clamping fixture (Fig. 188 Pos. 7) for damage 25/39 Instruction Manual WD290 © Belimed 10268...
  • Page 26: Servicing The Top Washing Arm

    Disassemble components (Fig. 108) Check the washing arm (Fig. 108 Pos. 2) for clogged nozzles and clean as required Check the rotor bearing (Fig. 108 Pos. 5) for wear and replace as required 26/39 Instruction Manual WD290 10268 © Belimed...
  • Page 27: Cleaning The Surface Sieve And Coarse Sieve

    Level float Coarse sieve Surface sieve Removing foreign bodies from the washing compartment Reinsert the coarse sieve correctly Reinsert the surface sieve correctly CAUTION! Never rinse without the surface sieve and coarse sieve! 27/39 Instruction Manual WD290 © Belimed 10268...
  • Page 28: Device Fails To Clean Properly

    Are the washing nozzles on the rack clogged? (See Chap. 8.1 "Loading racks") • Is the rack inserted correctly? (See Chap. 10.1 "Manual loading") • Have the items to be washed been proberly pre-treated? (See Chap. 4 "Pre- treating medical devices") 28/39 Instruction Manual WD290 10268 © Belimed...
  • Page 29: Faults

    Wait until it is enabled once again. load. Connection to PC for batch record- New program can be selected. No Communication ing is interrupted. Re-establish the connection Documentation system between the device and PC for batch recording. 29/39 Instruction Manual WD290 © Belimed 10268...
  • Page 30: Fault Text With Process Interruption

    Are the fuses intact? (inspection by the technical specialist) • Are the washing compartment doors correctly locked? (See Chap. 5 "Prepar- ing the device") • Fault code on the display? (See Chap. 16 "Faults") 30/39 Instruction Manual WD290 10268 © Belimed...
  • Page 31: Options

    Options 17 Options 17.1 Modem connection If a modem is installed, the device can be linked with the Belimed Service Centre. Inform the Service Centre before activating the modem (see Chap. 20 "Organi- sation Belimed AG". 17.1.1 Activating the connection...
  • Page 32: Tank Heating Manually Switchable

    (Fig. 231 Pos. 2) to steam heating (Fig. 231 Pos. 1) Restart program CAUTION! If the same fault still occurs after switching to steam, the technical specialist must be informed immediately (see Chap. 16.2 "Fault text with process inter- ruption") 32/39 Instruction Manual WD290 10268 © Belimed...
  • Page 33: Built-In Printer Cs / Us

    NOTE The function of the built-in printer is the same for both built-in versions CS/ Abb 232 Recess lid opens upwards Paper slit Costom built-in printer Paper feed button Print button (not active) 33/39 Instruction Manual WD290 © Belimed 10268...
  • Page 34: Operation

    If a problem occurs during a process, implement the following steps without de- lay: Press the emergency stop (Fig. 233) The ongoing process is interrupted immediately Rectify fault (see Chap. 16 "Faults") Restart program 34/39 Instruction Manual WD290 10268 © Belimed...
  • Page 35: Conformity And Certifications

    Conformity and certifications 18 Conformity and certifications 35/39 Instruction Manual WD290 © Belimed 10268...
  • Page 36: Glossary

    Washing Removal of contamination from the surface to be washed with an aqueous me- dium with or without process chemicals depending on requirements. 36/39 Instruction Manual WD290 10268 © Belimed...
  • Page 37 Washing chamber The part of the washing/disinfection device in which the load is treated. Note: Steam generator, piping, e.g. drains and connections, from which the chamber is separated, are not included. Warm water 37/39 Instruction Manual WD290 © Belimed 10268...
  • Page 38: Organisation Belimed Ag

    Organisation Belimed AG 20 Organisation Belimed AG 20.1 Manufacturer Belimed AG, Dorfstrasse 4, CH-6275 Ballwil (Switzerland) Phone +41 41 449 78 88 Fax +41 41 449 78 89 info@belimed.ch 20.2 Subsidiaries, Customer Service Austria Germany Branch Office West Belimed GmbH...
  • Page 39: Notes

    Notes 21 Notes 39/39 Instruction Manual WD290 © Belimed 10268...