Device enclosure

Cleaning
Clean the device enclosure using an alcohol based cleaning and disinfection wipe. ResMed has
validated CaviWipes1.
1. Wipe the exterior of the device using a wipe until visually clean following the manufacturer's
instruction for cleaning. Use a minimum of two wipes.
If visual debris is still present, perform the following:
2. Clean the exterior of the device with a dry, soft bristle brush and wipe the exterior of the device
using a new cleaning and disinfection wipe following the manufacturer's instruction for cleaning.
Disinfection
Repeat the first step with a new wipe and follow the manufacturer's instructions for disinfection.
Note: Failure to clean the component as indicated may result in inadequate disinfection.
Drying
Allow sufficient time for the device to air dry completely.
Note: Drying is not required after cleaning if disinfection is continued immediately.
Inspection
Perform a visual inspection of the device casing. If any visible deterioration is apparent (cracking,
crazing etc) discontinue use and contact your care provider or your ResMed Service Centre.
Air tubing, outlet connector and HumidAir 11 tub
Cleaning
1. Make a solution of a mild alkaline anionic detergent and water
instructions. ResMed has validated:
• Alconox at 1% (10 g/L) in water
2. Soak all components for 5-10 minutes. Agitate the component in the cleaning solution to ensure
there are no air bubbles.
3. Clean the inside and outside of all components with a soft bristle brush while soaking in a detergent
solution. Pay particular attention to all crevices and cavities.
• Tubing (Standard, SlimLine and ClimateLineAir 11): 3 minutes of brushing
Note: A soft bristle tube/bottle brush is required to clean the inside of the tubing. Remove tubes
from the detergent solution to assist brushing.
• Outlet connector: 1 minute of brushing
• HumidAir 11 Cleanable tub: 2 minutes of brushing
4. Thoroughly rinse each component as follows: in 5 liters of water
component by immersing it. Rinse tubing for 30-60 seconds. Agitate the component in the rinsing
water to ensure there are no air bubbles.
5. Repeat the rinse procedure two additional times using fresh water
Note: Failure to clean the component as indicated may result in inadequate disinfection.
at 69.8ºF to 131ºF (21ºC to 55ºC)
1
4
as directed by the manufacturer's
1
at ≤ 140ºF (≤ 60ºC) for each
1
for a total of three rinses.
1