General summary
ResMed has validated the following number of cycles for cleaning and disinfection using a manual
cleaning method.
Components
System component
Outlet connector
HumidAir 11 Cleanable tub
ClimateLineAir 11
Standard tubing
SlimLine
Device
Device Enclosure
Disassembling
Before disassembly, turn off the device and ensure the AC adaptor has been removed.
Cleanable Humidifier tub
1. Hold the humidifier tub at the top and bottom, press it gently and pull it away from the device.
Note: take care when handling the humidifier tub as the humidifier tub may be hot. Allow 10
minutes for the heater plate and any excess water to cool.
2. Hold the base of the humidifier tub and fully open the humidifier tub lid and pull it away so that it
easily detaches from the base.
Cleaning - Mild alkaline, anionic detergent (eg, Alconox™)
Chemical high level disinfection
eg, CIDEX
®
-OPA
120
120
30
30
30
Cleaning and low level disinfection
CaviWipes1™
2
Thermal high level disinfection
167ºF (75ºC) for 30 minutes
80
80
30
30
30