Download Print this page

Installing Sma Connectors; 34946A And 34947A Simplified Schematics - Agilent Technologies 34946A User Manual

Microwave switch/driver modules

Advertisement

Example: Reading the cycle count for a relay The following command reads back
the number of completed cycles for the channel 201 relay of a module installed
in slot 6.
DIAGnostic:RELay:CYCLes? (@6201)
Example: Clearing the cycle count for a relay The following command resets the
cycle count on channels 201 and 202 for a module in slot 1.
DIAGnostic:RELay:CYCLes:CLEar (@1201,1202)
Example: Resetting Module(s) to Power-On State The following command resets
a module in slot 4 to its power-on state.
SYSTem:CPON 4
Example: Enabling Verification The following command enables verification on
channels 201 and 202 for a module in slot 1. When verification is enabled, the
actual hardware state of each relay is sensed for the
correct state.
ROUTe:CHANnel:VERify:ENABle ON,(@1201,1202)

Installing SMA Connectors

When installing SMA connectors, it is recommend that you tighten them to 0.8
- 1.1 Nm (7-10 in-lbs) of torque.

34946A and 34947A Simplified Schematics

Figures 3 and 4 show the channel configurations for the 34946A and 34947A
modules, respectively.
50Ω
Agilent 34946A-34947A Microwave Switch/Driver Modules User's Guide
101
COM
102
Figure 3. 34946A Simplified Schematic.
34946A and 34947A Dual/Triple Microwave Switch Modules
50Ω
50Ω
201
50Ω
COM
202
5