3.3 Calibration Screen

Calibrate the voltages, temperatures, and currents to provide good gauging performance. Press the Calibration
button to select the Advanced Calibration window. See
3.3.1 Voltage Calibration
Measure the voltage from BAT+ to BAT– and enter this value in the Applied Voltage field and select the
Calibrate Voltage box.
Press the Calibrate Gas Gauge button to calibrate the voltage measurement system.
Deselect the Calibrate Voltage boxes after voltage calibration has completed.
3.3.2 Temperature Calibration
Enter the room temperature in the Applied Temperature field and select the Calibrate Temperature box and
select the thermistor to be calibrated. The temperature value must be entered in degrees Celsius.
Press the Calibrate Gas Gauge button to calibrate the temperature measurement system.
Deselect the Calibrate Temperature box after temperature calibration has completed.