Fluke 1523 Technical Manual page 88

Reference thermometer
Hide thumbs Also See for 1523:
Table of Contents

Advertisement

Find Quality Products Online at:
2. Set the UUT password enable by sending the following commands through the
serial port:
3. Set the UUT to the L75_OHMS cal range by sending the following commands
through the serial port:
4. Query the Ref points by sending the following Command:
5. Query the ADJ values for the range by sending the following Command:
6. Set the UUT in the LO_OHMS cal range by sending the following commands
through the serial port:
7. Query the Ref points by sending the following Command:
8. Query the ADJ values for the range by sending the following Command:
9. Set the UUT in the MED_OHMS cal range by sending the following commands
through the serial port:
10. Query the Ref points by sending the following Command:
11. Query the ADJ values for the range by sending the following Command:
12. Set the UUT in the HI_OHMS cal range by sending the following commands
through the serial port:
13. Query the Ref points by sending the following Command:
14. Query the ADJ values for the range by sending the following Command:
GlobalTestSupply
www.
Calibration of Your Reference Thermometer Readout
.com
Manual Calibration Process
83