Invacare Platinum POC1-100B Service Manual page 25

Oxygen concentrator
Hide thumbs Also See for Platinum POC1-100B:
Table of Contents

Advertisement

General
Alarm
Alarm Code Name
Alarm
Code
Name
Low Battery
0x7207 LPA_lowBattery30MinLeft
(~30
minutes
left)
Low Battery
0x2200 LPA_lowBattery15MinLeft
(~15
minutes
left)
1187375-B
All manuals and user guides at all-guides.com
Display Message
The display shows:
"Low Battery Charge
Unit"
The display shows:
"Low Battery Charge
Now"
Possible Reason
Solution
1. The concentrator is
1. Supply unit with
running off battery
power and the
battery charge
level is calculated
to only be able to
2. If alarm persists,
provide between
15 to 30 minutes
of additional
operation before
reaching 0% state
of charge.
1. The concentrator
1. Supply unit with
is running off
battery power
and the battery
charge level is
calculated to only
2. If alarm persists,
be able to provide
15 minutes or
less of additional
operation before
reaching 0% state
of charge.
Troubleshooting
external power
(AC/DC) or install
a charged battery
into the unit.
contact Invacare
for service.
external power
(AC/DC) or install
a charged battery
into the unit.
contact Invacare
for service.
25