Dc Current (Jaw); Ac Current (Jaw) - Fluke 377 Calibration Manual

Clamp meter
Hide thumbs Also See for 377:
Table of Contents

Advertisement

377/377 FC/378/378 FC
Calibration Manual

DC Current (Jaw)

To verify dc current measurement with the jaw:
1. Connect the normal output HI and LO of the calibrator to the HI and LO input of the
5500A/COIL.
2. Place the DUT inside the center of the 50-turn coil. Calibration accuracy to
specifications is guaranteed only when proper clamp alignment is made. The DUT
should be centered as much as possible on the base during verification.
3. Set the DUT to
4. Set the calibrator to the current:
a. Verify that the DUT reads within the display limits shown.
b. Verify the DUT reading for the remaining steps for dc current.

AC Current (Jaw)

To verify ac current measurement with the jaw:
1. Connect the normal output HI and LO of the calibrator to the HI and LO input of the
5500A/COIL.
2. Place the DUT inside the center of the 50-turn coil. Calibration accuracy to
specifications is guaranteed only when proper alignment is made. The DUT should be
centered as much as possible on the base during verification.
3. Set the DUT to
4. Set the calibrator to the current and frequency:
a. Verify that the DUT reads within the display limits.
b. Verify the DUT reading for the remaining steps for ac current.
4
Y
(DC current measurement mode).
s
(AC current measurement mode).

Advertisement

Table of Contents
loading

This manual is also suitable for:

377 fc378378 fc

Table of Contents