P03 - Capacitor Bank And Charging Test - Siemens MOBILETT Plus Service Instructions Manual

Hide thumbs Also See for MOBILETT Plus:
Table of Contents

Advertisement

Service programs
If an error is detected, the test program will be aborted, an audible beep will be signaled
and a message will be displayed:
kV
P02 Err 6 = Rotation error
If the test ends successfully, the display shows:
kV
P02 PASS = Rotation test passed in program P02

P03 - Capacitor bank and charging test

(Wiring diagram X037E, page 130)
When the test is started the display will show:
kV
CAP tESt = Capacitor test
During the test, the capacitor voltage is displayed except when the capacitance is shown.
The capacitor will first be charged to 350 V, then discharged to 229 V.
kV
CAP 350 = 350 Volt
The voltage will be held for 5 seconds before it is restored to 350 V. During the test, the
voltage in the capacitor bank is indicated, except when the capacitance is displayed:
kV
CAP 0.31 F = Capacitance 0.31 Farad
If an error is detected, the test program will be aborted. An audible beep will be signaled
and a message will be displayed:
kV
P03 Err"xx" = See actual "xx" in separate displayed message list
If the test ends successfully, the display will show:
kV
P03 PASS
Duration: Approx. 1.5 minutes.