1.6.3 NIBP Calibration Check

Revision A
DINAMAP® PRO Series 100-400V2 Monitor Pre-Service & Calibration Manual
This procedure verifies the linearity and calibration for both pressure
transducers (PT1 and PT2) over the range 0-250 mmHg.
To verify calibration it is necessary to have the following test
equipment:
- 2-tube NIBP hose, 12ft, p/n 107365
- DINAMAP Adult BP cuff p/n 2774 (or equivalent)
- Calibration Kit p/n 320246 with a manual inflation bulb
- NIST calibrated single-tube manometer
1. Set up the unit and calibration equipment as shown in Figure 4-1.
2. Power on the PRO Monitor and select More...
3. Select the Service mode and input 2 2 1 3.
4. From the Service menu, select the NIBP button.
5. From the NIBP menu, select the Check button.
2018554-001
Figure 4-1
17