Tone Alert / %Hu; Number Of Tubes; Ots; Wallstand Location - GE Precision 500D Operator's Manual

Table of Contents

Advertisement

3-4-9 HU Tone Alert / %HU

3-4-10 Number of Tubes

3-4-11 OTS

3-4-12 Wallstand Location

3-4-13 Fluoro Rates

3-4-14 Spot Rates

3-4-15 Max Record rate

3-4-16 System Log On

ILLUSTRATION 5-55
SYSTEM LOG-IN
om 5184516-100 Rev. 5
This is the Heat Units Remaining level at which the tone is sounded. This is only
followed when VA Site is set to "yes."
(1,2) When set to "1," only the under table tube is available, all references to Tube
2 are removed.
(Yes, No) When set to "yes," table bucky and table top are available however, the
number of tubes must be set to 2.
When set to "0", wall bucky will not be an option as a receptor.
(Continuous = 30, Pulsed = 15, 7.5, 4) These are set values and are options
depending if Pulsed Fluoro is set to yes or no.
(Single, 1, 2, 3, 4, 5, 6, 7.5) Available record rates.
(7.5) Fastest record rate
(Yes, No) When set to "yes" the system will require user to log on. If set to "no" the
Log out button is not displayed on the utilities screen.
5-57