Omron BP7350CAN Instruction Manual

Omron BP7350CAN Instruction Manual

Upper arm blood pressure monitor

Advertisement

EN
FR
7 Series Upper Arm / Au Bras Série 7
Blood Pressure Monitor /
Tensiomètre
BP7350CAN
Model / Modèle
Instruction Manual
/
Guide de l'utilisateur
3787149-9A

Advertisement

Table of Contents
loading

Summary of Contents for Omron BP7350CAN

  • Page 1 7 Series Upper Arm / Au Bras Série 7 Blood Pressure Monitor / Tensiomètre BP7350CAN Model / Modèle Instruction Manual Guide de l’utilisateur 3787149-9A...
  • Page 2: Table Of Contents

    Table of Contents Introduction ..........2 9. Optional Accessories ......38 Safety Instructions .......2 Using the AC Adapter (optional accessory) ......... 39 Intended Use ........2 Receiving and Inspection .....2 10. Specifications ........40 Symbols Glossary ........2 11. FCC/ISED Statement ......44 Important Safety Information ....3 12.
  • Page 3: Introduction

    Introduction Thank you for purchasing the OMRON BP7350CAN Blood Pressure Monitor. This monitor can be used by multiple users. This blood pressure monitor uses the oscillometric method of blood pressure measurement. This means this monitor detects your blood movement through your brachial artery and converts the movements into a digital reading.
  • Page 4: Important Safety Information

    Important Safety Information Read the Important Safety Information in this instruction manual before using this monitor. Follow this instruction manual thoroughly for your safety. Keep for future reference. For specific information about your own blood pressure, CONSULT WITH YOUR PHYSICIAN. Indicates a potentially hazardous situation which, if not Warning avoided, could result in death or serious injury.
  • Page 5 Important Safety Information AC Adapter (optional accessory) Handling and Usage • DO NOT use the AC adapter if this monitor or the AC adapter cable is damaged. If this monitor or the cable is damaged, turn off the power and unplug the AC adapter immediately.
  • Page 6 • Ensure that this monitor has acclimated to room temperature before taking a measurement. Taking a measurement after an extreme temperature change could lead to an inaccurate reading. OMRON recommends waiting for approximately 2 hours for the monitor to warm up or cool down when the...
  • Page 7 Important Safety Information AC Adapter (optional accessory) Handling and Usage • Fully insert the AC adapter into the outlet. • When unplugging the AC adapter from the outlet, be sure to safely pull from the AC adapter. DO NOT pull from the AC adapter cable. •...
  • Page 8: Know Your Monitor

    1. Know Your Monitor Contents/Product Includes Monitor, Easy Wrap ComFit Cuff, 4 “AA” batteries, instruction manual, quick start guide Monitor and Arm Cuff Front of the monitor Display [Memory] button [Forward/Backward] buttons [User ID selection] switch [START/STOP] button Left and right side of the monitor Air jack AC adapter jack (for optional AC adapter)
  • Page 9 1. Know Your Monitor Back and bottom of the monitor [Connection] button (Date/Time) Battery compartment Arm cuff (Upper arm circumference: 9" - 17" (22 - 42 cm)) Air plug Air tube...
  • Page 10: Display And Symbols

    1. Know Your Monitor Display and Symbols Average value symbol Bluetooth ON / OFF symbol Memory symbol Sync symbol Systolic blood pressure Low battery symbol / reading Depleted battery symbol Comparison display Date / Time display Pulse display / Memory High symbol number Diastolic blood pressure...
  • Page 11 1. Know Your Monitor Average value symbol Flashes when viewing the average of the latest 2 or 3 readings taken within the most recent 10 minute timeframe. Memory symbol Appears when viewing readings stored in the memory. Systolic blood pressure reading Comparison display Displays past readings.
  • Page 12 1. Know Your Monitor Sync symbol Flashes/appears when your data needs to be transferred because the stored memory is either almost, or completely full. Once you pair your monitor with your smart device, transfer your readings immediately before the monitor deletes the oldest reading.
  • Page 13 1. Know Your Monitor Irregular heartbeat symbol Appears when irregular rhythm is detected 2 or more times during a measurement. The symbol with the number appears as below: Irregular heartbeat is detected 2 times: Irregular heartbeat is detected 3 times or more: An irregular heartbeat rhythm is defined as a rhythm that is 25 % less or 25 % more than the average rhythm detected while your monitor is measuring...
  • Page 14: Classification Of Bp (Blood Pressure)

    1. Know Your Monitor Classification of BP (Blood Pressure) ACC/AHA 2017 Hypertension guidelines define categories of BP in adults as below. Categories of BP in Adults* BP Category Systolic BP Diastolic BP Normal < 120 mmHg < 80 mmHg Elevated 120 - 129 mmHg and <...
  • Page 15: Preparation

    2. Preparation Installing Batteries 1. Push down on the hook of the battery cover and pull downward. 2. Insert 4 “AA” batteries as indicated in the battery compartment. 3. Close the battery cover. As soon as batteries are correctly installed, the year flashes on the display.
  • Page 16: Setting The Date And Time

    •Set your monitor to the correct date and time before taking a measurement for the first time. •The date and time will automatically be set when the monitor is paired with “OMRON connect US/CAN” app. Refer to sub-section 4.1. •As soon as batteries are correctly installed, the year flashes on the display of your monitor.
  • Page 17 2. Preparation 3. Press the button to confirm the year, then the month will flash. Repeat the same steps to adjust the month, day, hour, and minute. Year Month Hour Minute 4. Press the button to store the setting. Your monitor automatically turns off 3 seconds after the minutes have been set in this sequence.
  • Page 18: Blood Pressure Measurement Tips

    2. Preparation Blood Pressure Measurement Tips To help ensure an accurate reading, follow these directions: •Stress raises blood pressure. Avoid taking measurements during stressful times. •Measurements should be taken in a quiet place. •It is important to take measurements at the same times each day. Taking measurements in the morning and in the evening is recommended.
  • Page 19: Applying The Arm Cuff

    •The blood pressure can differ between the right arm and the left arm, and the measured blood pressure readings can be different. OMRON recommends to always use the same arm for measurement. If the readings between both arms differ substantially, check with your physician to determine as to which arm to use for your measurements.
  • Page 20 2. Preparation 2. Apply the arm cuff to your left upper arm. Note • The bottom edge of the arm cuff should be 0.5 inch (1 to 2 cm) above the inside elbow. The air tube is on the inside of your arm and aligned with your middle finger.
  • Page 21: Sitting Correctly

    2. Preparation Sitting Correctly To take a measurement, you need to be relaxed and comfortably seated in a room with a comfortable temperature. Place your arm on the table. •Sit in a comfortable chair with your back and arm supported. •Keep your feet flat and your legs uncrossed.
  • Page 22: Using Your Monitor Without A Smart Device

    3. Using Your Monitor without a Smart Device Taking a Measurement You can use your monitor without pairing to a smart device. To pair your monitor with a smart device, refer to sub-section 4.1. Note •To stop the measurement, press the [START/STOP] button once to deflate the arm cuff.
  • Page 23 3. Using Your Monitor without a Smart Device 3. Press the [START/STOP] button to turn your monitor off. Note • Your monitor automatically turns off after 3 minutes. • Wait 2 - 3 minutes between measurements. The wait time allows the arteries to decompress and return to their pre- measurement state.
  • Page 24 3. Using Your Monitor without a Smart Device  If your systolic pressure is more than 210 mmHg If your systolic pressure is more than 210 mmHg, take a measurement manually. After the arm cuff starts to inflate, press and hold the [START/STOP] button until the monitor inflates 30 to 40 mmHg higher than your expected systolic pressure.
  • Page 25: Using Your Monitor With A Smart Device

    1. Enable Bluetooth on your smart device. 2. Download and install the “OMRON connect US/CAN” app onto your smart device. If you already have the “OMRON connect US/CAN” app and have created your account, open the app and add your new monitor.
  • Page 26: Taking A Measurement

    “OMRON connect US/CAN” app. • Your monitor automatically turns off after 2 minutes. • Be aware that OMRON will not be responsible for the loss of data and/or information in the app. • “OMRON connect US/CAN” is the only app that we recommend to use with your monitor to transfer data correctly.
  • Page 27: Disabling/Enabling Bluetooth

    4. Using Your Monitor with a Smart Device Disabling/Enabling Bluetooth Warning • This product emits radio frequencies (RF) in the 2.4 GHz band. DO NOT use this product in locations where RF is restricted, such as on an aircraft or in hospitals.
  • Page 28: Using Memory Function

    5. Using Memory Function Your monitor automatically stores up to 60 readings per user. Note •This monitor can be used by multiple users. Refer to sub-section 2.6 to switch user. Make sure which user ID stores your own readings. Viewing the Readings Stored in Memory 1.
  • Page 29: Viewing The Average Reading

    5. Using Memory Function Viewing the Average Reading Your monitor can calculate an average reading based on the latest 2 or 3 readings taken within the most recent 10 minute timeframe. Note •Set your monitor to the correct date and time before taking a measurement to view the average readings.
  • Page 30: Other Settings

    6. Other Settings Deleting All Readings Stored in Memory 1. Select your user ID. 2. Press the button. After a memory number appears, the latest reading will appear with a past reading. 3. While holding the button down, press and hold the [START/STOP] button for more than 3 seconds.
  • Page 31: Restoring Your Monitor To The Default Settings

    6. Other Settings 5. Press the [START/STOP] button to turn your monitor off. Note • All readings will be deleted. You cannot partially delete the readings stored in the memory. • Your monitor automatically turns off after 2 minutes. Restoring Your Monitor to the Default Settings To delete all the information stored in your monitor, follow the instructions below.
  • Page 32 6. Other Settings 3. Press the [START/STOP] button to turn your monitor off. Note • After the completion, the year flashes on the display. Refer to sub-section 2.2 to set the date and time manually, or refer to sub-section 4.1 to pair your monitor to set the date and time automatically.
  • Page 33: Error Messages And Troubleshooting

    7. Error Messages and Troubleshooting If any of the below problems occur during measurement, check to make sure that no other electrical device is within 12 inches (30 cm). If the problem persists, refer to the table below. Display/ Possible Cause Solution Problem Air plug is not...
  • Page 34 7. Error Messages and Troubleshooting Display/ Possible Cause Solution Problem Do not touch the arm cuff and/or bend the air tube while taking a The arm cuff is inflated exceeding the maximum measurement. allowable pressure. If inflating the arm cuff appears manually, refer to the end of sub-section 3.1.
  • Page 35 7. Error Messages and Troubleshooting Display/ Possible Cause Solution Problem Follow the instructions shown in the “OMRON connect US/CAN” app. If The monitor cannot connect to a smart the “Err” symbol still device or transmit data appears after checking correctly.
  • Page 36 Review sub-sections 2.3 - 2.5 and section 3. Follow the instructions shown in the smart device, Any other or visit the “Support” section in the “OMRON communication connect US/CAN” app for further help. If the issue occurs.
  • Page 37: Maintenance

    8. Maintenance Maintenance To protect your monitor from damage, follow the directions below: •Changes or modifications not approved by the manufacturer will void the user warranty. Caution • DO NOT disassemble or attempt to repair this monitor or other components. This may cause an inaccurate reading.
  • Page 38: Cleaning

    8. Maintenance Cleaning •Do not use any abrasive or volatile cleaners. •Use a soft dry cloth or a soft cloth moistened with mild (neutral) detergent to clean your monitor and arm cuff and then wipe them with a dry cloth. •Do not wash or immerse your monitor and arm cuff or other components in water.
  • Page 39: Optional Accessories

    9. Optional Accessories Easy Wrap ComFit Cuff AC Adapter Arm circumference 9" - 17" (22 - 42 cm) Model: HEM-FL31-B HEM-FL31-BZ Model: HEM-ADPTW5 Note •Do not throw the air plug away. The air plug can be applicable to the optional cuff. Protective LCD Cover For further information, visit our website at OmronHealthcare.com...
  • Page 40: Using The Ac Adapter (Optional Accessory)

    9. Optional Accessories Using the AC Adapter (optional accessory) Note •Make sure not to place your monitor in a location where it is difficult to plug and unplug the AC adapter. •We recommend keeping batteries in your monitor at all times, even if you choose to use the AC adapter.
  • Page 41: Specifications

    10. Specifications Model BP7350CAN HEM-7340T-CA Display LCD digital display Cuff pressure range 0 to 299 mmHg Blood pressure SYS: 60 to 260 mmHg measurement range DIA: 40 to 215 mmHg Pulse measurement 40 to 180 beats / min. range Pressure: ±3 mmHg Accuracy Pulse: ±5 % of display reading...
  • Page 42 10. Specifications Monitor: Approximately 16.2 oz. (460 g) not Weight including batteries Arm cuff: Approximately 5.8 oz. (163 g) Monitor: Approximately 7.5" (w) × 3.1" (h) × 4.1" (l) (191 mm × 79 mm × 105 mm) Dimensions Arm cuff: Approximately 5.7" × 20.9" (air tube: 29.5") (145 mm ×...
  • Page 43 10. Specifications About a wireless communication interference This product operates in an unlicensed ISM band at 2.4 GHz. In the event this product is used near other wireless devices such as microwave and wireless LAN, which operate on the same frequency band as this product, there is a possibility that interference may occur.
  • Page 44 10. Specifications Pictogram on the Product Types of marks used to indicate the correct positioning of the artery for the upper arm Identifies the type of cuff that is compatible with the blood pressure monitor...
  • Page 45: Fcc/Ised Statement

    11. FCC/ISED Statement FCC CAUTION Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment. This device complies with part 15 of FCC Rules and Innovation, Science and Economic Development Canada’s licence-exempt RSS(s).
  • Page 46 11. FCC/ISED Statement This equipment complies with FCC/ISED radiation exposure limits set forth for an uncontrolled environment and meets the FCC radio frequency (RF) Exposure Guidelines and RSS-102 of the ISED radio frequency (RF) Exposure rules as this equipment has very low levels of RF energy.
  • Page 47: Limited Warranty

    SOME STATES DO NOT ALLOW LIMITATIONS ON HOW LONG AN IMPLIED WARRANTY LASTS, SO THE ABOVE LIMITATION MAY NOT APPLY TO YOU. OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER SPECIAL, INCIDENTAL, CONSEQUENTIAL OR INDIRECT COSTS, EXPENSES OR DAMAGES.
  • Page 48 12. Limited Warranty This warranty provides you with specific legal rights, and you may have other rights that vary by jurisdiction. Because of special local requirements, some of the above limitations and exclusions may not apply to you. FOR CUSTOMER SERVICE Visit our web site at: OmronHealthcare.ca Call toll free:...