Download Print this page

Dismounting - Siemens MAMMOMAT 3000 Repair Instructions

Advertisement

Procedure

Dismounting

1.
2.
3.
4.
5.
NOTICE
6.
1
Fig. 1
Position of the protective conductors and the cable guide
Siemens-Elema AB
Solna, Sweden
Check that the swivel arm is in position 0°.
If not;
Switch the MAMMOMAT ON and turn the swivel arm to 0°.
Switch the MAMMOMAT OFF.
Secure optimum stability of the stand.
Using the clinometer, measure the inclination of the object table and note this
value (accuracy +1i).
Inclination x:_____i
Remove the shaft covers from swivel arm and tube.
Before removing the compression unit it is essential to mark up
the position of the base beam of the compression unit. This is to
ensure proper alignment when remounting the compression unit.
Remove the compression unit:
6.1. Remove cable guide (1/Fig. 1).
6.2. Remove earth screw (2/Fig. 1) and ground cables (3/Fig. 1).
2
3
Inclination y:_____i
SPB7-230.091.10
Rev. 01
02.00
2
Page 1 of 4
MAMMOMAT 3000
SPS-UD
2 - 1
2
Service

Advertisement

loading