The Humidity Level can be set to: Off or between 1 and 8, where 1 is the lowest humidity setting and
8 is the highest humidity setting.
For each humidifier setting, the Climate Control system delivers a constant amount of water vapor, or
absolute humidity (AH), to the patient's upper airway.
Automatic Adjustment
The humidifier and ClimateLineAir 11 heated air tubing are controlled by the Climate Control algorithm to
deliver constant humidity and temperature outputs. The system adjusts automatically to changes in:
ambient room temperature and humidity values
flow due to pressure changes
flow due to mask or mouth leak.