Connecting channels 2 to 4

Connect wires
1. Remove cap and ferrule from cable gland and slide onto cable.
2. Push cable through open gland and cable path.
3. Restore ferrule and tighten cap to lightly hold cable in place.
4. Fold signal cable screen back over outer sheath and ground beneath cable clamp.
In case of shielded cables, use metal cable glands for proper connection.
5. Connect wires to terminals using screwdriver.
6. Tighten cable gland.
The numbers in the graphics below refer to table Terminal layout (industrial enclosure)
(Page 52).
If connected as input or output - Industrial enclosure
4.6.6

Connecting the power supply

1. Open power supply terminal protection cover.
2. Remove cap and ferrule from cable gland and slide onto cable.
SITRANS FS230
Compact Operating Instructions, 05/2021, A5E38755607-AC
12
11
Active configuration
IO[3] (common)
IO[3]+ (active)
Functional Earth
4.6 Connecting industrial enclosure
13
12
Passive configuration
IO[3]- (passive)
IO[3] (common)
Functional Earth