Connecting
4.5 Connecting the external DSL
6. Connect the two, three or four wires to four-terminal block (Ch5-RTD or Ch6-RTD)
circuit terminals as required.
7. Assemble and tighten cable gland.
4.5.5

Connecting the SSL cable

M12 connector version
The SSL cable is provided with M12 style stainless steel plugs.
The cable shield is physically and electrically terminated within the body of the plug.
Note
Never pull the cable by the plug - only by the cable itself.
1. Connect DSL using the supplied 4-wire cable with M12 connectors to both the external DSL
and the transmitter.
Note
Grounding
The SSL cable shield is only mechanically connected to the grounding terminal (PE), when
the M12 plug is correctly tightened.
48
2-wire device
Short
Isource +
Vsense +
Vsense -
Isource -
3-wire device
Short
Isource +
Vsense +
Vsense -
Isource -
Compact Operating Instructions, 05/2021, A5E38755607-AC
. Short-
4-wire device
Isource +
Vsense +
Vsense -
Isource -
SITRANS FS230