Ni 6711/6713-12-Bit Resolution; Ni 6722/6723-13-Bit Resolution - National Instruments NI-DAQ mx Calibration Procedure

Ao waveform
Table of Contents

Advertisement

NI 6711/6713—12-Bit Resolution
Range (V)
Minimum
Maximum
–10
10
–10
10
–10
10
Set Point (MHz)
NI 6722/6723—13-Bit Resolution
Range (V)
Minimum
Maximum
–10
10
–10
10
–10
10
Set Point (MHz)
© National Instruments Corporation
Table 4. NI 6711/6713 Analog Output Values
Test Point
Location
Value (V)
0
0.0
Pos FS
9.9900000
Neg FS
–9.9900000
Table 5. NI 6711/6713 Counter Values
Upper Limit (MHz)
5
Table 6. NI 6722/6723 Analog Output Values
Test Point
Location
Value (V)
0
0.0
Pos FS
9.9000000
Neg FS
–9.9000000
Table 7. NI 6722/6723 Counter Values
Upper Limit (MHz)
5
24-Hour Ranges
Lower
Upper
Limit (V)
Limit (V)
–0.0059300
0.0059300
9.9822988
9.9977012
–9.9977012
–9.9822988
Lower Limit (MHz)
5.0005
24-Hour Ranges
Lower
Upper
Limit (V)
Limit (V)
–0.0070095
0.0070095
9.8896747
9.9103253
–9.9103253
–9.8896747
Lower Limit (MHz)
5.0005
15
AO Waveform Calibration Procedure for NI-DAQmx
1-Year Ranges
Lower
Upper
Limit (V)
Limit (V)
–0.0059300
0.0059300
9.9818792
9.9981208
–9.9981208
–9.9818792
4.9995
1-Year Ranges
Lower
Upper
Limit (V)
Limit (V)
–0.0070095
0.0070095
9.8892582
9.9107418
–9.9107418
–9.8892582
4.9995