Lock; Apply Scaling; Assigned; Available - Omega SP-016 User Manual

Heat flux thermopile smart probe
Table of Contents

Advertisement

5.3.3.1 Lock

If set, the user-specified units of measure string (4-character maximum) will be used in place of the default DIN.

5.3.3.2 Apply Scaling

If set, the user-defined Offset and Gain values will be used to adjust the sensor reading. For more information on
Gain and Offset, refer to the Smart Sensor Manual.

5.3.3.3 Assigned

The Assigned bit will always read as 0. Refer to the Smart Sensor Device Interface documentation for further
information.

5.3.3.4 Available

The Available bit will always read as 0. Refer to the Smart Sensor Device Interface documentation for further
information.

5.3.4 DIO Device Configuration

The DIO Device Configuration allows enabling each of the 2 input bits and selecting whether the input is active HIGH
or active LOW. If the Invert Bit is set the signal will be Active Low.
7
0

5.3.4.1 Invert

If the Invert bit is set the input is active LOW.

5.3.4.2 Enable

If the Enable bit is set the input is enabled.

5.3.5 DIO IPSO Definition

The DIO input IPSO definition provides signal range, measured min/max values, IPSO object type information.
Offset
0x00
0x02
0x04
0x08
0x0c
0x10
0x14
Result = (Raw Reading * Gain) + Offset
DIO Device Configuration
6
5
Reserved
0
0
Name
Sensor Type
Precision
Sensor Trigger Function
Min Measured
Max Measured
Min Range
Max Range
4
3
DIN 1
ENABLE
INVERT
0
1
Value
3349
Bit Mapped Digital
0
Provides reading of xxx
??
See section Sensor Trigger Function
??
Minimum reading since last reset
??
Maximum reading since last reset
0
Minimum reading
3
Maximum reading
2
1
DIN 0
ENABLE
INVERT
1
1
Description
0
1
23 |
P a g e