• Avoid placing the instrument near equipment that radiates heat.
Do not place the instrument near gas burners, electric heaters or
ovens.
• Do not place the instrument near equipment that generates in-
tense magnetic fields such as electric welding equipment, high
frequency furnaces, pole transformers, etc.
• Protect the instrument from excessive dust.
• Connect the instrument to power lines that are free from sudden
changes or voltage fluctuations.
• If you must use power motor driven equipment (such as a stirrer
or shaker) in the same line as your instrument, ensure that a
noise reduction unit is in the same power line.

3.2 Conditions for the installation

Check the following requirements before installing the instrument:
Power supply and ground
Line voltage:
115 or 230VAC (make sure the voltage selector
switch is correctly set).
Frequency:
50 or 60 Hz
Power capacity: 180 W
Ground
A grounded outlet should be within 2 m of the in-
terminal:
strument.
Gas supply for flushing the monochromator (optional)
Nitrogen
Required pressure 0.5 bar (7 psi)
Gas consumption 6 L/min

3.3 Removal of the shipping protection

Open the Scanner 4 by sliding the door to the right side and remove
the knurled screws and the three brackets (1) (2) (3). Keep the
whole shipping protection set in a suitable place. Prior to shipping
the instrument the brackets must be screwed back on.
Fig. 2: The shipping protection
Instruction Manual, June 2018
TLC Scanner 4
(1)
(2)
(3)
Page 7