Maintenance Data Sheet - CAMAG TLC SCANNER 4 Instruction Manual

Table of Contents

Advertisement

CAMAG Maintenance data sheet
TLC Scanner 3/4
April 2012/UB
Purpose
The maintenance data sheet informs about maintenance interval of the respective instru-
ment as well as the proposal for IQ/OQ interval if applicable. In addition, it identifies con-
sumable parts with the respective replacement cycle.
Maintenance interval
Mainte-
12 Months
nance
IQ/OQ
12 Months
Consumable parts
Part No.
Description
027.6440
Hg lamp
027.6441
D2 lamp
027.6442
W lamp
352.0014
Compartment illumination UV254
115.0123
L-shaped magnetic strip
115.0124
Magnetic strip
* Latest when the intensity is dropping under 60%

5.6 Maintenance data sheet

Instruction Manual, June 2018
TLC Scanner 4
Replacement cycle
Upon need*
Upon need*
Upon need*
Upon need
Upon need
Upon need