M4735A Heartstream XL Defibrillator/Monitor
Delivery of the shock is confirmed by the message 6KRFN 'HOLYHUHG and the
shock counter is updated.
)LJXUH  6KRFN 'HOLYHUHG 'LVSOD\
HR
LEAD
SpO2
ALARM
SELECT
ON/OFF
Pads
120
Shock Delivered

Automatic Re-analysis On

If Automatic Re-analysis is on, the Heartstream XL analyzes the ECG fol-
lowing delivery of the shock. You are prompted to press
tional shock is advised. This cycle repeats until the rhythm converts or a
shock series is complete. (A shock series may be configured to 2, 3, or 4
shocks.)

Automatic Re-analysis Off

If Automatic Re-analysis is off, the Heartstream XL monitors the ECG for
potentially shockable rhythms provided Rhythm Monitoring is on and
prompts you to press
without being prompted by pressing
Shocks: 1
00:00:40
$1$/<=(
if one is detected. You can initiate analysis
$1$/<=(
.
Defibrillating
3
6+2&.
, if an addi-
3-13