Step 2: Ecg Analysis And Monitoring - Agilent Technologies M3860A User Manual

Semi-automatic external defibrillator aed
Hide thumbs Also See for M3860A:
Table of Contents

Advertisement

M3860A and M3861A H
EARTSTREAM
Step 2:
ECG Analysis and Monitoring
Follow the instructions provided by the H
and screen prompts in the order indicated.
As soon as the FR2 detects that the defibrillation pads are connected properly,
it automatically begins analyzing the patient's heart rhythm. The
H
FR2 model M3860A also displays the patient's ECG on the
EARTSTREAM
screen.
If no shock is advised, the H
prompts to tell you so. The FR2 instructs you to perform CPR if needed, and
performs background monitoring of the patient's ECG while you give
appropriate care to the patient.
No more voice prompts are given, until and unless the H
detects a change in the patient's heart rhythm that may be a shockable
rhythm, or
detects interference with rhythm analysis.
If the H
FR2 detects a potentially shockable heart rhythm, it
EARTSTREAM
automatically goes back to analyzing the rhythm to see if a shock is advised.
If a shock is advised, the H
delivery. It gives the voice warnings and screen prompts to tell you that a
shock is advised, Make sure no one is touching the patient. While the
H
FR2 is charging, it continues to analyze the patient's heart
EARTSTREAM
rhythm. If the rhythm changes and a shock is no longer appropriate, the
H
FR2 disarms and dumps the charge. Voice and display
EARTSTREAM
prompts advise you what action to take.
NOTE: When the H
EARTSTREAM
by pressing the On/Off button to turn off the FR2 and return it to standby mode. (See
the Defibrillator discussion in Appendix B, Technical Specifications, for details on
disarming the FR2.)
FR2 Defibrillator

Step 2: ECG Analysis and Monitoring

EARTSTREAM
FR2 provides voice and screen
EARTSTREAM
FR2 charges to prepare for shock
EARTSTREAM
FR2 is fully charged, you can disarm it at any time
FR2 voice
FR2:
EARTSTREAM
3-3