Gain and Offset

The values in Tables 6 and 7 do not include DAQ device errors.
Note

SCXI-1102/B/C Front and Rear Panel Diagrams

© National Instruments
Table 6 contains the test limits to use when verifying and adjusting the gain
and offset of the SCXI-1102 and SCXI-1102B. Table 7 contains the test
limits to use when verifying and adjusting the gain and offset of the
SCXI-1102C. If the SCXI-1102/B/C was calibrated within the last year, the
test point value should fall between the Lower Limit and Upper Limit
values found in Tables 6 or 7.
Table 6. SCXI-1102 and SCXI-1102 B Test Limits
Gain
Test Point (V)
1
9.800000
0.000000
–9.800000
100
0.098000
0.000000
–0.098000
Gain
Test Point (V)
1
9.800000
0.000000
–9.800000
100
0.098000
0.000000
–0.098000
Table 8 shows the pin assignments for the SCXI-1102/B/C front panel
connector. Table 9 shows the pin assignments for the SCXI-1102/B/C rear
signal connector.
Upper Limit (V)
9.80345
0.001002
–9.79655
0.098060
0.000030
–0.097940
Table 7. SCXI-1102 C Test Limits
Upper Limit (V)
9.80345
0.001002
–9.79655
0.098060
0.000030
–0.097940
15
Lower Limit (V)
9.79655
-0.001002
–9.80345
0.097940
–0.000030
–0.098060
Lower Limit (V)
9.79655
-0.001002
–9.80345
0.097940
–0.000030
–0.098060
SCXI-1102/B/C Calibration Procedure