Electromagnetic Flowmeter Converter
1)
Supply power, see Instrument Tag.
2)
Supply cable, e.g. shielded 2x1.5 mm
3)
External zero return, passive, over contact with internal voltage supply
terminals U2 (+24 V DC) and 22, switch S902 closed.
External totalizer reset, passive over contact with internal voltage supply
terminals U2 (+24 V DC) and 31, switch S902 closed.
4)
External totalizer reset with external voltage supply (for galv.isolation) terminals G2 (–)
and 31 (+24 V DC). Switch S902 on the connection board must be opened.
External zero return with external voltage supply (for galv.isolation) terminals G2 (–)
and 31 (+24 V DC). Switch S902 on the connection board must be opened.
Dynamic zero compensation with external voltage supply (for galv.isolation) terminals G2 (–)
and A1 (+24 V DC). Switch S902 opened. The fluid must be at absolute zero flowrate
the metering tube must be completely filled, see 3.3.
Scaled pulse output, 24 V DC active, load ≥150
5) *
terminals V1, V2; Function 9, 11 forward
terminals V3, V4; Function 9, 11R reverse
Scaled pulse output, passive, relay contact, ≤28 V DC, ≤250 mA, ≤3 W, fmax 20 Hz
6) *
terminals V1, V2; Function 55, 56 forward
terminals V3, V4; Function 57, 58 reverse
7)
Option, data link RS 485. terminals V1, V2, V3, V4; Function T-, T+, R-, R+.
8)
Option, data link Profibus DP, terminals V1, V2, V4, G2, Function RxD/TxD-P(B),
RxD/TxD-N(A), +5V (VP), DGND(G2) terminals V4, G2 only to be used for bus termination.
Alarm max., relay contact ≤28 V DC, ≤250 mA, ≤3 W,
9)
passive or optocoupler U
Alarm min., relay contact ≤28 V DC, ≤250 mA, ≤3 W,
10)
passive or optocoupler U
Forward/reverse direction signal, relay contact ≤28 V DC, ≤250 mA, ≤3 W. terminals 44, 45, 46.
11)
12)
Current output, R
Ω
≤2000
R
L
13) a) Alarm output, relay contact opens at alarm, ≤28 V DC, ≤250 mA, ≤3 W, terminals V5, V6; Function 39, 40 or
b) Alarm output, optocoupler, U
If a passive scaled pulse output is required in addition to the data link, then the alarm output is not available.
c) Scaled pulse output, passive, relay contact (closes) ≤28 V DC, ≤250 mA, ≤3 W
terminals V5, V6; Function 55, 56 forward.
14)
Shielded signal cable (10 m included with shipment from ABB), ABB Part No. D173D018U02.
15)
Supply voltage for flowmeter primaries with preamplifiers (1/25" - 5/16" [DN 1 - DN 8] always include a preamplifier),
terminals +U, -U.
*) Scaled pulse output, active or passive, pulse width can be set from 0.032 to 2000 ms.
Terminal Designations 1/25" to 16" [DN 1 to DN 400]
8
2
. terminals M1/M3.
≤25 F, I
≤7.5 mA. terminals P1, P2; Function E7, C7.
CE
CE
≤25 F, I
≤7.5 mA. terminals P3, P4; Function E8, C8.
CE
CE
Ω
≤550
0/4–20 mA, 0–10–20 mA, 4–12–20 mA, R
L
0–5 mA. terminals –, +.
≤25 F, I
≤7.5 mA. terminals V5, V6; Function E9, C9.
CE
CE
Ω
, fmax 10 kHz
≤1000
L
Ω
0/2–10 mA,