Examples of Application Combinations

Checkweighing with statistical evaluation
You want to check a piece count, and have the results that lie within the tolerance range statistically evaluated and printed as a ISO/GLP-com-
pliant record.
Settings (changes in the factory settings required for this example):
Setup: Application parameters: Application 1: Counting
Setup: Application parameters: Application 2: Checkweighing
Setup: Application parameters: Application 3: Statistics: Automatic storage: On, first value at stability
Setup: Application parameters: Application 3: Statistics: Source of data for auto storage: Application 2
Setup: Application parameters: Application 3: Statistics: Evaluated value: Calculated
Setup: Application parameters: Application 3: Statistics: Evaluation mode, MR function: Intermediate evaluation, display+print
Setup: Application parameters: Printout configuration: ISO/GLP/GMP printout: Always
Step
1. If necessary: turn on the balance
and enter the settings given above
2. Place reference sample quantity
on the balance
3. Initialize the balance
4. Remove reference sample quantity
5. Initialize Checkweighing
Toggle to Checkweighing
Key (or instruction)
e
Place parts on the balance
Start
soft key
Unload the balance
d
Display/Output
--------------------
18.03.1997
SARTORIUS
Mod.
Ser.-no.
Ver.-no.
ID
---------------------
C-ID
nRef
wRef
21.03500 g
n
Qnt
+
09:41
LA4200S
60419914
01-35-18
10 pcs
1
10 pcs
111