Invacare Carroll CS Series User Manual page 57

Medical bed
Table of Contents

Advertisement

Effect
Actuator not working.
1153410-Q
Possible cause
HI/LO lock-out is on.
Loose wire connections.
Verification
For non-openbus (CS3, CS5)
- A Lockout box (a square
black box) with a toggle
switch. When the switch is
towards "ON" side, the HILO
on the hand pendant will not
function. There is no LED
light.
For openbus beds (CS7, CS9,
CS9FX600) - The lockout box
(a flatter rectangular white
box) with a toggle dome
button and LED light. When
the light is on, the HILO on
the hand pendant will not
function.
Lockout LED on the Attendant
Control Pad (ACP) is ON.
Check cable connections
visually.
Troubleshooting
Corrective action
Rotate the lockout dial to
the "unlocked" position
(horizontal), or flip switch to
OFF.
Toggle-off the individual
function lockout (LED should
be OFF)
Connect any loose cable
couplings and/or power
supplies.