Specifications; Electrical Characteristics - Fluke i310s Instruction Sheet

Ac/dc current clamp
Hide thumbs Also See for i310s:
Table of Contents

Advertisement

CAT III equipment is designed to protect against
the transients in the equipment in fixed
equipment installations, such as distribution
panels, feeders and short branch circuits, and
the lighting systems in large buildings.
Do not use Clamp in wet environments or in
locations that hazardous gases exist.

Specifications

Electrical Characteristics

All accuracies stated at 23°C ± 1°C (73.4 °F ± 33.8 °F)
Current Range
Inrush Current
Output Sensitivity
Accuracy
Bandwidth to Meet
Accuracy
Specification
Phase Shift below 1
kHz
Resolution
Load impedance
Conductor Position
Sensitivity
Frequency Range
(small signal)
Temperature
Coefficient
Power Supply
Working Voltage
Battery Life
30 A and 300 A ac rms or
± 45 A and 450 A dc
600 A ac rms Max
10 mV/A (30 A)
1 mV/A (300 A)
(30 A range) ± 1 % of reading ± 50 mA
(300 A range) ± 1 % of reading ± 300 mA
@25 °C, Bandwidth dc to 1 kHz
1 kHz
< 2 degrees
± 50 mA (30 A)
± 100 mA (300 A)
> 10 k Ω and ≤ 100 pF
± 1.5 % relative to center reading
DC to 20 kHz (-3 dB)
± 0.01 % of reading / °C
9 V Alkaline, NEDA 1604/PP3 IEC 6LR61
300 V ac rms or dc
30 hours, low battery indicator