Specifications - Hach CL17 Instrument Manual

Chlorine analyzer
Hide thumbs Also See for CL17:
Table of Contents

Advertisement

Specifications

General
Sample Requirements
Reagent/Standard Requirements
Electrical
Page 6
Specifications
Specifications are subject to change without notice.
Display: LCD, 3½-digit measurement readout and six-character
alphanumeric scrolling text line.
Enclosure: IP62-rated with the gasketed door latched
Instrument Dimensions: 34.3 cm width x 41.9 cm height x 19.1 cm depth
(13.5 in. x 16.5 in. x 7.5 in.)
Mounting: Wall mount
Instrument Shipping Weight: 11.3 kg (25 lb)
Sample Flow Rate to Sample Conditioning: 200 to 500 mL/min
Inlet Pressure to Instrument:
1 to 5 psig; 1.5 psig is optimum. Exceeding 5 psig can cause sample
tubing failure.
Inlet Pressure to Sample Conditioning: 1.5 to 75 psig (with sample tube
level with the bottom of the instrument—see
Sample Temperature Range: 5 to 40 °C (41 to 104 °F)
Inlet Fitting: at instrument—¼-inch OD polyethylene tubing with
quick-disconnect fitting
Drain Fitting: ½-inch hose barb
Sample Conditioning: Use provided sample conditioning
Maximum Reagent Usage: one-half liter per month (each of the two
reagents)
Reagent Containers: High-density polyethylene (2) ½-liter bottles.
Reagent Containment: Reagent bottles are contained inside the analyzer
enclosure and are vented externally.
Power Requirements: 100–115/230 V ac (selection switch inside
instrument); 95 VA, 50/60 Hz, 2.5 Amp fuse
Power Connection: Connection made by three wire barrier terminal block
through a ½-inch conduit hole in the case. Wire range: 12–18 AWG.
Installation Category: II
Figure 6 on page
19)